| Catalog Number |
168-800 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
04/02/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The reason for reoperation is deflation.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.This is a known potential adverse event addressed in the product labeling.
|
| |
|
Event or Problem Description
|
|
Patient reported a right side deflation.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified: mold green in the surface of the shell, opening and crease flat.Leak test was performed and found opening on posterior.A microscopic analysis was performed which identify a sharp edge opening assessed as unidentified tear opening.A dimension measurement in the shell was performed which identify the thickness within specification.The fill test inspection was performed, the result is no blockage.Based on the device analysis the final assessment is: a sharp opening on posterior due to an unidentified (tear) opening.
|
| |
|
Event or Problem Description
|
|
The device has been explanted.
|
| |
|
Search Alerts/Recalls
|