• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) STYLE 168 SALINE FILLED BREAST IMPLANT; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) STYLE 168 SALINE FILLED BREAST IMPLANT; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE Back to Search Results
Catalog Number 168-800
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 04/02/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.The reason for reoperation is deflation.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.This is a known potential adverse event addressed in the product labeling.
 
Event or Problem Description
Patient reported a right side deflation.Device remains implanted.
 
Additional Manufacturer Narrative
Device evaluation: visual analysis of the returned device identified: mold green in the surface of the shell, opening and crease flat.Leak test was performed and found opening on posterior.A microscopic analysis was performed which identify a sharp edge opening assessed as unidentified tear opening.A dimension measurement in the shell was performed which identify the thickness within specification.The fill test inspection was performed, the result is no blockage.Based on the device analysis the final assessment is: a sharp opening on posterior due to an unidentified (tear) opening.
 
Event or Problem Description
The device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 168 SALINE FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
CS 
MDR Report Key8554540
Report Number9617229-2019-02768
Device Sequence Number13945152
Product Code FWM
UDI-Device Identifier10888628000940
UDI-Public10888628000940
Combination Product (Y/N)N
PMA/510(K) Number
P990074
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/29/2021
Device Catalogue Number168-800
Device Lot Number3110754
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/15/2019
Initial Date Received by Manufacturer 04/05/2019
Supplement Date Received by Manufacturer06/03/2019
Initial Report FDA Received Date04/26/2019
Supplement Report FDA Received Date06/03/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age51 YR
-
-