• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ENDOSCOPIC CANN DRILL BIT 7MM; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ENDOSCOPIC CANN DRILL BIT 7MM; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT Back to Search Results
Catalog Number 013660
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/28/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported a rust on the drill bit.No patient injuries or delay were reported as the failure was found after the surgery.
 
Additional Manufacturer Narrative
Device has not been returned for examination, however a photograph was supplied.Review of the photograph confirmed the reported rusting like oxidation.The drills are water stained and contaminated with what appears to be human matter likely causing the observed oxidation.The most common cause of rust formation on stainless steel is improper cleaning.If the product is returned in the future the complaint can be reopened and evaluated.Initial reporter occupation: physician.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPIC CANN DRILL BIT 7MM
Common Device Name
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
MDR Report Key8554845
Report Number1219602-2019-00471
Device Sequence Number1586470
Product Code HWE
UDI-Device Identifier03596010091611
UDI-Public03596010091611
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number013660
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/28/2019
Supplement Date Received by Manufacturer06/14/2019
Initial Report FDA Received Date04/26/2019
Supplement Report FDA Received Date06/18/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-