• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MFG DC BOULDER VALLEYLAB; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG DC BOULDER VALLEYLAB; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number FORCEARGONII-20
Medical Device Problem Codes Energy Output Problem (1431); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Blood Loss (2597); No Code Available (3191)
Date of Event 03/15/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during the liver mass biopsy, the unit's hand pieces would not fire properly - it was very weak and never arced properly.Had it turned all the way up and both tanks were full.Had to open the patient (converted to laparotomy) to control bleeding.Three (3) laparoscopic tips and two (2) hand pieces were opened ¿ could not get any of them to work properly.Another generator was used with a regular hand pencil cautery because had to open the patient.Used cautery and endo ties to complete the surgery.There was a 6000 ml of bleeding and had to transfuse plasma x5, platelets x 2, rbc x 15.Patient was discharged.
 
Additional Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.It was reported that the unit would not fire.The reported issue could not be confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALLEYLAB
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
MDR Report Key8556110
Report Number1717344-2019-00559
Device Sequence Number19397347
Product Code GEI
UDI-Device Identifier10884524002552
UDI-Public10884524002552
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K964636
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFORCEARGONII-20
Device Catalogue NumberFORCEARGONII-20
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/21/2019
Initial Date Received by Manufacturer 04/08/2019
Supplement Date Received by Manufacturer04/08/2019
Initial Report FDA Received Date04/26/2019
Supplement Report FDA Received Date01/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age52 YR
Patient Weight110
-
-