It was reported that the patient was implanted with vns to treat depression, and that the patient experienced a cardiac arrest at the end of a surgery in november 2018.Per the reporter, the patient¿s battery has been dead for some amount of years.There was no indication the surgery was associated with vns; however, a concern was expressed that the cardiac arrest may be associated with the lead remaining attached to the vagus nerve.No additional or relevant information has been received to date.
|
Further follow up with the anesthesiologist confirmed that the vns device was recently ruled out as a possible contributor to the cardiac event, as it was confirmed to be "non-functional," which was defined as the generator battery being completely depleted.It was further confirmed that the vagus nerve was not manipulated during the procedure, and that the patient has no evidence of cardiac disease.It was mentioned that the patient will have the occasional unifocal pvc and fainting spells attributed to "vasovagal syndrome." additionally, it was noted that the patient does have a history of possible cataplexy but was never medicated for it.No other relevant information has been received to date.
|