• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH VIPER PRIME CFXFN XTB 7X45MM S; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH VIPER PRIME CFXFN XTB 7X45MM S; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Model Number 186770445S
Medical Device Problem Code Disconnection (1171)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/01/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Device is available for evaluation.Investigation will be conducted.Follow up will be filed with the investigation results.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the viper prime screw saddle dislodged from screw tulip.Removed and new screw inserted.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Udi: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Visual examination revealed signs of operative use as evidenced by superficial markings, with damage to the drive feature of the screw head and the inner cap (saddle) is dislodged by 90 degrees and fell apart from its intended location.It was also noted the tip of screw was broken.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.The root cause of the screw falling apart cannot be determined from the sample and the information provided.A potential root cause may be excessive force inadvertently placed on the drive feature of the screw during tightening, potentially at a significant angle, resulting in the saddle falling apart during use.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Udi: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIPER PRIME CFXFN XTB 7X45MM S
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key8557337
Report Number1526439-2019-51572
Device Sequence Number2948203
Product Code NKB
UDI-Device Identifier10705034530253
UDI-Public(01)10705034530253
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K173095
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number186770445S
Device Catalogue Number186770445S
Device Lot NumberTBVZC
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/07/2019
Initial Date Received by Manufacturer 04/02/2019
Supplement Date Received by Manufacturer05/07/2019
06/20/2019
Initial Report FDA Received Date04/26/2019
Supplement Report FDA Received Date05/30/2019
06/20/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-