| Model Number |
S401 |
| Medical Device Problem Codes |
Over-Sensing (1438); Pacing Problem (1439)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/12/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this pacemaker system exhibited oversensing on the right atrial (ra) lead, which resulted in pacing inhibition.Subsequently, a revision procedure was performed.The pacemaker was explanted and replaced and the ra lead was surgically abandoned and replaced.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction to field d7: explant date.Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the device was performed.A visual inspection of the device header and case noted no anomalies.The device was then exposed to simulated heart load conditions, and the pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported oversensing and pacing inhibition.
|
| |
|
Event or Problem Description
|
|
This supplemental report is being filed as the device evaluation was completed.
|
| |
|