| Catalog Number |
140.686 |
| Medical Device Problem Code |
Failure to Osseointegrate (1863)
|
| Health Effect - Clinical Code |
Tissue Damage (2104)
|
| Date of Event |
08/25/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device was sent to the manufacturer in a manner unsuitable for investigation.Exemption number: e2015015.Instradent usa, inc is submitting the report on behalf of neodent - jjgc.
|
| |
|
Event or Problem Description
|
|
The clinician reported that over 1 month after the immediate dental implant was placed in ada site 5 in the patient's mouth, the implant did not achieve osseointegration in type ii bone.The clinician noted the patient's bruxism.The implant was torqued manually to 45 ncm.There were no reported patient complications.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was sent to the manufacturer in a manner unsuitable for investigation.Exemption number: e2015015.Instradent usa, inc is submitting the report on behalf of neodent - jjgc.Considering the surgical methodology, it was seen that the dentist has selected an implant design which is not recommended for the patient's bone quality.The information provided in this report was based on information given by the dentist in neodent¿s products warranty form that was recorded in the system that is used by both the manufacturer and the subsidiary.The original complaint and the product were received by the subsidiary and did not arrive in the manufacturer yet.When this happens, a new evaluation of the complaint will be carried out and, if necessary, a new follow-up report will be send.
|
| |
|
Event or Problem Description
|
|
The clinician reported that over 1 month after the immediate dental implant was placed in ada site 5 in the patient's mouth, the implant did not achieve osseointegration in type ii bone.The clinician noted the patient's bruxism.The implant was torqued manually to 45 ncm.There were no reported patient complications.
|
| |
|
Search Alerts/Recalls
|