| Model Number |
AC*T DIFF 2 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); Test Result (2695)
|
| Date of Event |
01/21/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The field service engineer (fse) was on site and confirmed what the customer reported.The fse observed that the cause of high platelets recoveries was a kinked eva tubing that prevented sweep flow to properly work.To resolve the issue the fse dressed the eva tubing to prevent it from kinking.The fse also replaced the vacuum isolator chamber (vic), the aspiration probe, the probe wipe, and the hgb lamp pn 6805344 as preventative maintenance.Bec internal identifier - (b)(4).
|
| |
|
Event or Problem Description
|
|
The customer reported high platelets (plt) results for patients samples on their coulter ac·t diff 2 analyzer.Neither patient results or raw data were provided by the customer.There were no erroneous patient¿s results released from lab nor effect to patient treatment.
|
| |
|