BECKMAN COULTER COULTER AC·T DIFF 2 ANALYZER; COUNTER, DIFFERENTIAL CELL Back to Search Results
Model Number AC*T DIFF 2
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problems No Consequences Or Impact To Patient (2199); Test Result (2695)
Event Date 01/21/2019
Event Type  malfunction  
Manufacturer Narrative
The field service engineer (fse) was on site and confirmed what the customer reported.The fse observed that the cause of high platelets recoveries was a kinked eva tubing that prevented sweep flow to properly work.To resolve the issue the fse dressed the eva tubing to prevent it from kinking.The fse also replaced the vacuum isolator chamber (vic), the aspiration probe, the probe wipe, and the hgb lamp pn 6805344 as preventative maintenance.Bec internal identifier - (b)(4).
 
Event Description
The customer reported high platelets (plt) results for patients samples on their coulter ac·t diff 2 analyzer.Neither patient results or raw data were provided by the customer.There were no erroneous patient¿s results released from lab nor effect to patient treatment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COULTER AC·T DIFF 2 ANALYZER
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 s. kraemer blvd.
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th avenue
miami FL 33196 2031
Manufacturer Contact
laurie o'riordan
11800 sw 147th avenue
miami, FL 33196-2031
3053802874
MDR Report Key8558391
MDR Text Key143393034
Report Number1061932-2019-01541
Device Sequence Number1
Product Code GKZ
UDI-Device Identifier15099590264796
UDI-Public(01)15099590264796(11)NO-DATA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K990352
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/26/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAC*T DIFF 2
Device Catalogue Number6605500
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/21/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/01/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-