| Model Number |
HOYA ISERT 251 (+17.50 D) |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
11/11/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted after the initial 30-day reporting period deadline; it is part of a corrective action being implemented per internal document capa-(b)(4) for outside us like products reporting.This initial and final emdr is being submitted to fda for outside us like products reporting.A portion of the device was returned to the manufacturer and a product investigation was conducted.The results are listed below: the lens was ejected out from the injector tip and was not returned.Therefore, visual inspection could not perform and further information could not provide about the lens.The slider and the rod could advance fully.Trace of lubricant was found inside the injector tip.No abnormalities were found in production and inspection records of the product.Exact root cause is not determined.Risk analysis conducted on the product line and failure code is noted below: a review of the most recent complaint trending data indicates that no significant trends have been identified at this time, and no capa is required as part of the product evaluation.
|
| |
|
Event or Problem Description
|
|
When the lens was inserted in to the eye, a piece of the optic was missing from the outside edge.The lens had to be cut in half and removed, and another was then inserted.
|
| |
|