Catalog Number 71436384
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/05/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a tkr the device broke while it was being impacted.No backup device available.No injuries, a delay from 0-30 minutes was reported.
 
Additional Manufacturer Narrative
The associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.One of the condyles fractured off the device and all pieces were returned.The device was manufactured in 2017 and exhibits signs of moderate wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII PLASTIC REAMTHRU FEM TRL SZ 4 LT
Common Device Name
PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8559678
Report Number1020279-2019-01693
Device Sequence Number11310476
Product Code JWH
UDI-Device Identifier00885556625828
UDI-Public00885556625828
Combination Product (Y/N)N
Initial Reporter CountryPK
PMA/510(K) Number
K142807
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number71436384
Device Lot Number17LB00050
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/02/2019
Initial Date Received by Manufacturer 04/05/2019
Supplement Date Received by Manufacturer04/05/2019
Initial Report FDA Received Date04/28/2019
Supplement Report FDA Received Date07/21/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-