• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. TANDEM BIPOLAR COCR 46OD 28ID; PROSTHESIS, HIP, HEMI-, FEMORAL, METAL/POLYMER, CEMENTED OR UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. TANDEM BIPOLAR COCR 46OD 28ID; PROSTHESIS, HIP, HEMI-, FEMORAL, METAL/POLYMER, CEMENTED OR UNCEMENTED Back to Search Results
Catalog Number 71322046
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Injury (2348)
Date of Event 04/05/2019
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that patient had a dislocated bipolar hip, on (b)(6) 2019 surgeon tried to reduce the hip and head dissociated from the tandem bipolar shell.On (b)(6) 2019 a revision surgery was performed, implants were removed and replaced.
 
Additional Manufacturer Narrative
The associated complaint devices were not returned.A clinical evaluation was conducted and the provided x-rays confirm the reported dislocation; however, they do not provide a reason/root cause for the dislocation.Without the requested clinical relevant documents we are unable to conduct a thorough medical assessment.Therefore, no medical assessment is warranted at this time.This complaint will be re-evaluated if more information becomes available.A review of complaint history on the listed parts revealed no prior complaints for the listed batches with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TANDEM BIPOLAR COCR 46OD 28ID
Common Device Name
PROSTHESIS, HIP, HEMI-, FEMORAL, METAL/POLYMER, CEMENTED OR UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8559682
Report Number1020279-2019-01695
Device Sequence Number11310626
Product Code KWY
UDI-Device Identifier03596010492456
UDI-Public03596010492456
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K023743
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number71322046
Device Lot Number17DM13295
Initial Date Received by Manufacturer 04/05/2019
Supplement Date Received by Manufacturer10/28/2019
Initial Report FDA Received Date04/28/2019
Supplement Report FDA Received Date11/02/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-