| Catalog Number |
04.01.0165 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/01/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Preliminary investigation performed by r&d (b)(4): as reported by the sales representative, the insertion of the screw was totally carried out with the "glenoid polyaxial screwdriver - non-retentive" (ref.(b)(4)).The use of this instrument instead of the "polyaxial screwdriver - modular tip" (ref.(b)(4)), prevents the petals from being captured, exposing the head to an increased risk of breakage.Batch review performed on 29 apr 2019: lot 174752: (b)(4) items manufactured and released on 30-jan-2018.Expiration date: 2023-01-16.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.
|
| |
|
Event or Problem Description
|
|
During the primary surgery the head of the screw got damaged when the surgeon tried to insert it with the screwdriver.The screw was further damaged extracting it with a pincer.The bone was sclerotic.The surgery was completed successfully using a new screw.No critical delay during the surgery and no parts fell into the patient.The "glenoid polyaxial screwdriver - non-retentive" (ref.(b)(4)) was used to insert the screw.
|
| |
|
Additional Manufacturer Narrative
|
|
On may 09th 2019 we discovered that the implant reference reported in the initial report was not correct.The correct reference is 04.01.0165.
|
| |
|