| Catalog Number |
2C8931 |
| Medical Device Problem Code |
Failure to Prime (1492)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The single-use device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
This report summarizes 1 malfunction events.It was reported that a clearlink continu-flo solution set would not prime.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information : the single-use device was received for evaluation.A visual inspection was performed and noted that all components were correctly placed and according to specifications.Functional and pressure testing were performed and no blocks or leaks were identified.The reported issue was not verified.The device was found to operate per specification.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|