Catalog Number 2C8931
Medical Device Problem Code Failure to Prime (1492)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The single-use device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
This report summarizes 1 malfunction events.It was reported that a clearlink continu-flo solution set would not prime.There was no patient involvement.No additional information is available.
 
Additional Manufacturer Narrative
Additional information : the single-use device was received for evaluation.A visual inspection was performed and noted that all components were correctly placed and according to specifications.Functional and pressure testing were performed and no blocks or leaks were identified.The reported issue was not verified.The device was found to operate per specification.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK CONTINU-FLO MULTI-PORT MANIFOLD
Common Device Name
STOPCOCK, I.V. SET
Manufacturer (Section D)
BAXTER HEALTHCARE - DOMINICAN REPUBLIC
haina, san cristobal
MDR Report Key8560704
Report Number1416980-2019-02385
Device Sequence Number11347706
Product Code FMG
UDI-Device Identifier00085412071725
UDI-Public(01)00085412071725
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)Y
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device No Information
Device Catalogue Number2C8931
Device Lot NumberDR18G02076
Initial Date Received by Manufacturer 03/31/2019
Supplement Date Received by Manufacturer06/30/2019
Initial Report FDA Received Date04/29/2019
Supplement Report FDA Received Date07/29/2019
Patient Sequence Number1
-
-