|
On (b)(6) 2019, a helpline call reported a "msd not functional" alarm after a unilateral dvt procedure.The msd had gone in smoothly.The alarm was not resolved.Treatment proceeded with using the ekos catheter as a catheter-directed thrombolysis/with lytic only (cdt).The icu nurse reported that a stat-lock type of device was used to hold catheter in place.When fluids were hooked up, the system sounded a downstream occlusion alarm on the iv pump.The nurse had to remove the "stat-lock" to get fluids to run.The patient was reportedly fine and stable.On (b)(6), the territory manager reported that, during follow-up, a venogram showed that there was still thrombus in the common femoral down to the popliteal.The physician used angiojet which cleared the rest of the thrombus.The patient was taken back to icu in good condition.
|