The broaches are getting stuck in the me 1000 adapter.As per sales rep, he spoke with (b)(6) (event reporter), the reason that the broaches are getting stuck is a result of the broach trunnion being damaged while using the calcar planer while not fully seated over the broach trunnion.This is causing burrs on the broach trunnions.
|
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: the device associated with this report was not returned.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot null.Device history batch null.Device history review null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|