• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INSPIRE MEDICAL SYSTEMS INSPIRE MODEL 4323 RESPIRATORY SENSING LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INSPIRE MEDICAL SYSTEMS INSPIRE MODEL 4323 RESPIRATORY SENSING LEAD Back to Search Results
Model Number 4323
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Date of Event 04/05/2019
Type of Reportable Event Serious Injury
Event or Problem Description
The physician informed inspire of a sense lead revision that took place on (b)(6) 2019.Patient complained of pain.The respiratory lead was readjusted within the intercostal space and the system tested intraoperatively.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRE MODEL 4323 RESPIRATORY SENSING LEAD
Common Device Name
INSPIRE MODEL 4323 RESPIRATORY SENSING LEAD
Manufacturer (Section D)
INSPIRE MEDICAL SYSTEMS
5500 wayzata blvd.
suite 1600
golden valley MN 55416
MDR Report Key8560985
Report Number3007666314-2019-00032
Device Sequence Number1836875
Product Code MNQ
UDI-Device Identifier10855728005526
UDI-Public10855728005526
Combination Product (Y/N)Y
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P130008
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/29/2019
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number4323
Device Catalogue Number900-002-010
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/05/2019
Initial Report FDA Received Date04/29/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Date Report Sent to FDA01/01/2005
-
-