Model Number A22205C
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As part of our investigation, olympus followed-up with the user facility to obtain additional information.The user facility further reported that the loop at the distal end of the device broke off and fell into the patient¿s prostatic urethra.The loop was retrieved and the device was removed from the patient without harm.In addition, the procedure was prolonged by 5 minutes.There was no sparking/arcing observed during procedure.There was no unexpected bleeding to the patient observed and the patient did not require a longer stay or additional treatment.Prior to procedure, the electrode was inspected with no irregularities noted.The generator settings were set to 200 cut, 80 coag; no error message was displayed.The oem conducted a review of the device history records (dhr) for the referenced device/lot number and there was no deviations or non-conformities noted during the manufacturing process.The referenced device will not be returned to olympus, therefore, the exact cause of the reported event cannot be confirmed.However, based on similar reported complaints, the reported event can potentially occur if the electrode comes into contact (unintended) with other metal parts, e.G.Surgical instruments while the high-frequency output was activated.As a preventive measure, the instruction manual contains several warning and caution statements in an effort to prevent damage to the electrode.Visually inspect the instrument prior to use.Warped electrodes, defective insulation and broken, cracked, or irregular cutting loops represent a danger for both the patient and the surgeon and must not be used.If the instrument is damaged or does not function properly, replace it.
 
Event or Problem Description
Olympus was informed that during a transurethral resection of a prostate (turp) procedure, the loop at the distal end of the device appeared to be thinning out and the loop would break at the apex.The intended procedure was completed utilizing a second like device from the same lot.The reported event did not impact to the outcome of the procedure.No patient injury was reported.The reported device was discarded by the user facility following the procedure.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide additional information related to the event.Update to section b5.
 
Event or Problem Description
Information was provided to olympus which stated "the loop did not fall into the patient, the loop failed.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE, LOOP
Common Device Name
HF RESECTION ELECTRODE LOOP
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
MDR Report Key8561385
Report Number2951238-2019-00782
Device Sequence Number1537357
Product Code GCP
UDI-Device Identifier14042761036709
UDI-Public14042761036709
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K790071
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 03/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberA22205C
Device Catalogue NumberA22205C
Device Lot Number1000032107
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 04/25/2019
Supplement Date Received by Manufacturer05/14/2019
Initial Report FDA Received Date04/29/2019
Supplement Report FDA Received Date03/18/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-