Model Number 7010
Device Problem Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/26/2019
Event Type  malfunction  
Manufacturer Narrative
Patient identifier: (b)(6).Returned product consisted of a quantum maverick balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed multiple kinks along the hypotube.The hypotube is separated 50.3cm from the hub and appears to have been kinked prior to inflation.Microscopic examination revealed damage to the tip.There is contrast present in the inflation lumen and balloon.There is blood present in the guide wire lumen and the balloon is loosely folded.Inspection of the remainder of the device presented no other damage or irregularities.
 
Event Description
Reportable based on device analysis completed on 08-apr-2019.It was reported that shaft kink occurred.The 93% stenosed, 2.75x11mm target lesion was located in the moderately tortuous and moderately calcified left anterior descending artery.A 2.75mm x 12mm quantum maverick balloon catheter was advanced for dilatation.However, it was noted that the balloon delivery shaft was kinked at 55cm from the physician's hand.The procedure was completed with another of the same device.No patient complications were reported and patient status was stable.However, returned device analysis revealed shaft break.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTUM MAVERICK
Type of Device
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC SCIMED, INC
two scimed place
maple grove MN 55311
Manufacturer Contact
sonali arangil
two scimed place
maple grove, MN 55311
6515827403
MDR Report Key8561673
MDR Text Key143555037
Report Number2134265-2019-04514
Device Sequence Number1
Product Code LOX
UDI-Device Identifier08714729392668
UDI-Public08714729392668
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P860019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 04/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/12/2021
Device Model Number7010
Device Catalogue Number7010
Device Lot Number0022374913
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/12/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/08/2019
Initial Date FDA Received04/29/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/13/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age40 YR
Patient Weight74
-
-