ENCORE MEDICAL L.P. PS KNEE; INSERT, POSTERIOR STABILIZED, SERIES 500, SIZE6 9MM CM Back to Search Results
Model Number 360-09-506
Health Effect - Clinical Code Pain (1994)
Date of Event 04/01/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reason for this revision surgery was due to pain and instability.The previous surgery and the revision detailed in this investigation occurred over 9 years and 9 months apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported components used in the previous surgery met design and manufacturing requirements.There was an nonconformance associated with the part# 360-09-506, insert posterior stabilized size 6 9 millimeter which documents a nonconformance that out 20 quantity lot, 1 item was rejected and scrapped due to out of tolerance dimension.All other items in the lot were met the fit, design and functional requirements.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on- going issues that are in need of review.The root cause of this complaint was a revision surgery due to pain and instability.There were no findings during this investigation that indicate that the reported device was the source or had a direct connection with the patient's event.There are multiple factors that may contribute to the event that are outside the control of djo surgical such as: poor bone density, inadequate soft tissue support, soft tissue impingement, degenerative bone disease, excessive weight, unbalanced joints, patient bone deterioration, patient activities or trauma.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Event or Problem Description
Revision surgery - due to post operative posterior stabilized (ps) insert broken.Painful and unstable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PS KNEE
Common Device Name
INSERT, POSTERIOR STABILIZED, SERIES 500, SIZE6 9MM CM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key8561952
Report Number1644408-2019-00364
Device Sequence Number16692283
Product Code JWH
UDI-Device Identifier00888912116268
UDI-Public(01)00888912116268
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K933539
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2013
Device Model Number360-09-506
Device Catalogue Number360-09-506
Device Lot Number53978091
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/29/2019
Initial Report FDA Received Date04/29/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
-
-