• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS - HOUSTON PULSE GEN MODEL 103; GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS - HOUSTON PULSE GEN MODEL 103; GENERATOR Back to Search Results
Model Number 103
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Code Available (3191)
Event Date 10/14/2016
Event Type  Injury  
Event Description
The patient underwent battery replacement on (b)(6) 2016 due to reported battery depletion.Explant notes for the patient's generator explant in 2016 were received and indicated that the generator was flipped and turned and the wires were coiled in a twisted manner into a knot.This was able to be un-round and untied.After explant, it was stated the 106 was then secured more efficiently with silk sutures.A lead test was performed and showed good impedance 2309 ohms.While the notes don¿t indicate that the reason for surgery was due to the device migration (patient was referred due to low battery), the notes do state that this was a more difficult than usual procedure due to findings of the wires and generator twisting resulting in extra time to unwind and inspect the wires and electrode connector.No additional information has been received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 103
Type of Device
GENERATOR
Manufacturer (Section D)
CYBERONICS - HOUSTON
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS - HOUSTON
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
rachel kohn
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key8562074
MDR Text Key143461630
Report Number1644487-2019-00820
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 04/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date12/06/2012
Device Model Number103
Device Lot Number201689
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 04/04/2019
Initial Date FDA Received04/29/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/17/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age28 YR
-
-