ERIKA DE REYNOSA, S.A. DE C.V. LIBERTY CYCLER SET, SINGLE CONN./EXT. DL; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
|
Back to Search Results |
|
Catalog Number 050-87216 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 04/16/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
|
|
Event Description
|
It was reported that a peritoneal dialysis (pd) patient discovered a fluid leak on the cassette after ending their pd treatment.Fluid was also reported to be in/on the cycler.The patient reported receiving an air detected in cassette alarm during dwell 3 of 4 of treatment and the treatment was cancelled.It is unknown at which point in therapy the leak may have begun.The cause of the leak is unknown.The patient was advised to discontinue the use of their cycler and follow up with their peritoneal dialysis nurse (pdrn).A new cycler was issued to the patient.It was reported that an alternate treatment option was available, but the patient is unfamiliar with it.Upon follow up, the patient contact confirmed that the cycler was returned and replaced.All other information regarding the reported event was unknown.
|
|
Manufacturer Narrative
|
Plant investigation: as the device was not returned to the manufacturer, a physical evaluation could not be performed.A batch records review was conducted by the manufacturer for the reported lot.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.The entire lot has been sold and distributed.In addition, a device history review was performed and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The lot met all specifications for release.A product history review did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
|
|
Search Alerts/Recalls
|
|
|