Catalog Number 050-87216
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/16/2019
Event Type  malfunction  
Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Event Description
It was reported that a peritoneal dialysis (pd) patient discovered a fluid leak on the cassette after ending their pd treatment.Fluid was also reported to be in/on the cycler.The patient reported receiving an air detected in cassette alarm during dwell 3 of 4 of treatment and the treatment was cancelled.It is unknown at which point in therapy the leak may have begun.The cause of the leak is unknown.The patient was advised to discontinue the use of their cycler and follow up with their peritoneal dialysis nurse (pdrn).A new cycler was issued to the patient.It was reported that an alternate treatment option was available, but the patient is unfamiliar with it.Upon follow up, the patient contact confirmed that the cycler was returned and replaced.All other information regarding the reported event was unknown.
 
Manufacturer Narrative
Plant investigation: as the device was not returned to the manufacturer, a physical evaluation could not be performed.A batch records review was conducted by the manufacturer for the reported lot.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.The entire lot has been sold and distributed.In addition, a device history review was performed and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The lot met all specifications for release.A product history review did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIBERTY CYCLER SET, SINGLE CONN./EXT. DL
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
MDR Report Key8562180
MDR Text Key143824137
Report Number8030665-2019-00655
Device Sequence Number1
Product Code FKX
UDI-Device Identifier00840861100750
UDI-Public00840861100750
Combination Product (y/n)N
PMA/PMN Number
K043363
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 05/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number050-87216
Was Device Available for Evaluation? No
Device Age MO
Initial Date Manufacturer Received 04/16/2019
Initial Date FDA Received04/29/2019
Supplement Dates Manufacturer Received05/02/2019
Supplement Dates FDA Received05/07/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
DELFLEX PD FLUID; LIBERTY SELECT CYCLER
-
-