| Model Number |
S722 |
| Medical Device Problem Code |
Interrogation Problem (4017)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/15/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This device has been returned back from the field for analysis.This report will be updated upon completion of analysis.
|
| |
|
Event or Problem Description
|
|
It was reported that during an implant procedure, this pacemaker was connected to the leads however the field was unable to establish telemetry and interrogate it.Despite troubleshooting, the physician could not resolve the issue so the pacemaker was removed and replaced prior to pocket closure.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the device was performed.A visual inspection of the device header and case noted no anomalies.The device was then exposed to simulated heart load conditions, and the pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
|
| |
|
Event or Problem Description
|
|
It was reported that during an implant procedure, this pacemaker was connected to the leads however the field was unable to establish telemetry and interrogate it.Despite troubleshooting, the physician could not resolve the issue so the pacemaker was removed and replaced prior to pocket closure.No adverse patient effects were reported.
|
| |
|