|
Device evaluation summary: the device was returned to mentor without fluid inside.No foreign material was observed within the device or on the shell surface.Mentor product analysis lab discovered a rent at the patch junction of the anterior aspect.Microscopic examination was performed.No evidence of instrument damage was observed.No other anomalies were discovered.The complaint was confirmed since a rupture was found on the device.However, at this point it is not possible to determine the root cause of such damage.A manufacturing record evaluation (mre) of lot number 7396548 was reviewed, and no anomalies were found related to this complaint.In addition, the mre review verifies that the device was manufactured in accordance with documented specification and procedures.There is no evidence that the issue is related to manufacturing.Breast implants are not considered lifetime devices and deflation is a known complication associated with these devices as referenced in our product insert data sheet.Manufacturer¿s reference number: (b)(4).
|
|
It was reported that a (b)(6) female patient underwent primary breast reconstruction with a mentor artoura 650cc breast tissue expander that deflated postoperatively.On (b)(6) 2019, during surgery to remove the device, a hole was identified on the junction patch near the filling port.
|