| Catalog Number |
486611640 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Codes |
Pain (1994); Discomfort (2330); Injury (2348)
|
| Date of Event |
01/22/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Stryker was made aware of a related device complaint on (b)(6) 2019 (mrn 3004024955-2019-00004).On (b)(6) 2019, additional information was provided to stryker indicating a complaint related to a concomitantly implanted screw.Three catalog numbers were provided for the screw, 48611640, 486611645, and 486611650; it is unknown which catalog number pertains to this device.Catalog number 48611640 has been populated in this report.This device information will be updated in a supplemental report upon receipt of additional clarifying information from the reporting physician.
|
| |
|
Event or Problem Description
|
|
The device was implanted (b)(6) 2016.The patient began to experience discomfort attributed to the concomitantly implanted cage in (b)(6) 2017.Revision surgery was performed to address the concomitant device on (b)(6) 2019.During revision, the physician noted that this device (l4 screw) was, "slightly loose," then removed and replaced it.
|
| |
|
Event or Problem Description
|
|
The device was implanted (b)(6) 2016.The patient began to experience discomfort attributed to the concomitantly implanted cage in (b)(6) 2017.Revision surgery was performed to address the concomitant device on (b)(6) 2019.During revision, the physician noted that this device (l4 screw) was, "slightly loose," then removed and replaced it.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual, dimensional, functional and materials analysis could not be performed as the device was not returned.A review of complaint and device history records could not be performed as a valid lot code was not provided and could not be obtained.The updated information from the operation notes and review of primary surgery records states that one screw was "slightly loose." this cannot be confirmed based on images received.Not returned by hospital.
|
| |
|