| Catalog Number |
54750014530 |
| Medical Device Problem Codes |
Inadequacy of Device Shape and/or Size (1583); Malposition of Device (2616)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/14/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.X-ray review result: post-op ct show a thoracic construct with bi-cortical purchase of both screws at this level.Insertion distance and screw length is determined by the performing surgeon and no hardware failure is present in this image.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Pre operative diagnosis for this procedure: dislocation fracture at t5-t6 type of procedure or technique used: t3-t4 and t6-t7 percutaneous pedicle screw (pps) fusion it was reported that intra-op, due to unavailability of 4.5mm x 25mm screws, 4.5mm x 30mm screws were inserted forcibly using o-arm navigation.As a result, the screws on both sides of t3 protruded the anterior wall of the vertebral body, causing an event that hit the lung and the thoracic aorta.No patient complications were reported but it was nearly a catastrophe.Final tightening was performed forcibly with the construct.There was a delay of less than 60 minutes in the overall procedure time.No injury to the patient was reported.
|
| |
|