Catalog Number 101011020
Medical Device Problem Codes Loss of Osseointegration (2408); Device-Device Incompatibility (2919); Unintended Movement (3026)
Health Effect - Clinical Codes Inadequate Osseointegration (2646); No Code Available (3191)
Date of Event 04/09/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient's left hip was revised due to femoral component loosening at the bone to implant interface.The actis stem appeared to be undersized and collapsed into varus.Doi: (b)(6) 2018; dor: (b)(6) 2019; left hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Per internal procedures, the event information was reviewed.No investigational inputs were received.For this investigation, no immediate action was required as no alleged deficiency with the device was identified.No device associated with this report was received for examination.In order to determine if a lot related issue was possible, a worldwide complaint database search was performed.A worldwide complaint database search found no additional related reports against the provided product code/lot number combination.A search of the depuy nonconformance (nc) quality system found no nc¿s associated with this product/lot combination.Based on the inability to find any nc¿s against the provided product code/lot code combination, it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.Investigational inputs were requested as indicated per internal procedures for this failure mode, however, no information was provided to depuy without the physical complaint sample associated with this report, it was not possible to determine if the device failed to meet specifications at the time it was released for distribution.The device associated with this event was used in the treatment of the patient as prescribed by the presiding surgeon.From the event information received, it was not possible to determine the relationship of the device to the reported event.No evidence was found indicating product error was a contributing factor.The need for corrective action was not indicated.Depuy considers the investigation closed.Should additional info be received, the investigation will be re-opened as necessary.Device history lot
=
> hk5997.Device history review
=
> a search of the depuy nonconformance (nc) quality system found no nc¿s associated with this product/lot combination.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot: hk5997.Device history batch: null.Device history review: a search of the depuy nonconformance (nc) quality system found no nc¿s associated with this product/lot combination.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIS COLLARED STD SIZE 2
Common Device Name
HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork IN
EI 
MDR Report Key8564629
Report Number1818910-2019-91980
Device Sequence Number19422801
Product Code KWL
UDI-Device Identifier10603295380429
UDI-Public10603295380429
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number101011020
Device Lot NumberHK5997
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/09/2019
Supplement Date Received by Manufacturer05/16/2019
05/16/2019
Initial Report FDA Received Date04/30/2019
Supplement Report FDA Received Date05/16/2019
05/17/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age83 YR
Patient Weight47
-
-