| Model Number |
LF1723 |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
03/29/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Evaluation summary: the product was not returned; however, a picture was provided by the customer for analysis.The picture showed a device with a disengaged jaw pin.The reported condition was confirmed.Without the device, pmv could not determine what caused the reported issue of the complaint.The investigation could not determine the root cause of the customer's report based on the picture sample sent.The customer is advised to return the device to pmv, so a complete evaluation can be performed.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during anterior resection with ileostomy covering, the jaw of the instrument was unable to close and when inspected they saw a small pin on the sterile draping sheet outside the patient's abdominal cavity, and a new device was used.The patient had a surgical site infection due to surgery and dressing and debridement was done.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: one device was received for evaluation.The returned product did not meet specification as received.Visual inspection found the device's cam pin was disengaged from the jaw.The disengaged cam pin was returned.The reported condition was confirmed.The manufacturing engineer was notified and confirmed that the device's cam pin weld was broken.The broken weld cause the cam pin to become disengaged.The investigation identified the root cause of the reported event to be a manufacturing error.The device cam pin is laser welded and should not disengage.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|