The product sample was discarded by the customer; therefore, analysis of the product could not be conducted.Without a sample or product photo, root cause of reported issue and exact location of the reported leak was unable to be determined.Based on the information provided, issue appears to be a heat exchanger leak.There currently have been no change in complaint levels for hex leaks.3m will continue to monitor.End of report.
|
|
A hospital anesthesia technician reported a (b)(6) male patient, weighing 20kg, had liver transplant surgery on (b)(6) 2019.The patient was being administered rbc fluid through a 3m¿ ranger¿ standard flow disposable set (model 24200) warmed by a 3m¿ ranger¿ warming unit (model not specified).Prior to administration, the fluid warming cassette was loaded into the warming unit.The warming unit was turned off at this time.The patient was administered rbc fluid for approximately 2 to 2 ½ hours.No pressure device or pump was used.No manual pressure was applied to the fluid bags.The employee alleged, approximately 2 to 2 ½ hours after the start of administration, the anesthesiologist discovered blood was leaking from the ranger ¿ unit.The anesthesiologist took action to prevent any bodily harm to the patient.The employee reported the warming unit was unable to be evaluated at the hospital due to blood contamination.
|