| Catalog Number |
365965 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/15/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that bd microtainer® tubes with lh (lithium heparin) were clotting.No serious injury or medical intervention was reported.The following information was provided by the initial reporter: material no.365965, batch no.830556n.It was reported that the tubes were clotting.The tubes were clotting.Date of event: (b)(6) 2019 (2 of 2).Dates of event: (b)(6) 2019: no patient identifiers, unknown lot, no samples, (b)(6) 2019: female, (b)(6) yrs old, lot 830556n, samples available.
|
| |
|
Event or Problem Description
|
|
It was reported that bd microtainer® tubes with lh (lithium heparin) were clotting.No serious injury or medical intervention was reported.The following information was provided by the initial reporter: material no.365965, batch no.830556n.It was reported that the tubes were clotting.The tubes were clotting.Date of event: (b)(6) 2019 (2 of 2).Dates of event: (b)(6) 2019: no patient identifiers, unknown lot, no samples.(b)(6) 2019: female, 18 yrs old, lot 830556n, samples available.
|
| |
|
Additional Manufacturer Narrative
|
|
H.6.Investigation: investigation summary: bd received samples from the customer facility for investigation.The samples were evaluated and the customer's indicated failure mode for clotting with the incident lot was not observed.Additionally, retention samples were selected from bd inventory for evaluation/testing and upon completion, no issues were observed relating to clotting as all samples met specifications.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Investigation conclusion: based on evaluation of the customer and retain samples, the customer¿s indicated failure mode for clotting with the incident lot was not observed as all samples met the required specifications.Root cause description: based on the investigation, a root cause could not be determined.The product was found to be in conformance and meet release specifications.Rationale: complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.
|
| |
|