BD CARIBE LTD. BD MICROTAINER® TUBES WITH LH (LITHIUM HEPARIN); BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 365965
Medical Device Problem Code Coagulation in Device or Device Ingredient (1096)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/07/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medical device expiration date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
 
Event or Problem Description
It was reported that sample collected in the bd microtainer® tube with lh (lithium heparin) clotted.This complaint was created to capture 1 of 2 related incidents.The following information was provided by the initial reporter: "the tubes were clotting.".
 
Additional Manufacturer Narrative
Investigation summary: bd had not received samples or photos from the customer facility for evaluation; therefore, the investigation was limited.Additionally, bd was unable to determine the specific lot number associated with this complaint.Therefore, a review of the device history record could not be conducted.Investigation conclusion: as no samples or photos were received for evaluation, the customer's indicated failure mode was not observed by bd.Root cause description: as there was no sample or photo available for evaluation, a root cause could not be determined.Rationale: complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
It was reported that sample collected in the bd microtainer® tube with lh (lithium heparin) clotted.This complaint was created to capture 1 of 2 related incidents.The following information was provided by the initial reporter: "the tubes were clotting.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MICROTAINER® TUBES WITH LH (LITHIUM HEPARIN)
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BD CARIBE LTD.
road 31
k.m. 24.3
juncos
MDR Report Key8565546
Report Number2618282-2019-00164
Device Sequence Number1809421
Product Code JKA
UDI-Device Identifier50382903659659
UDI-Public50382903659659
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K991702
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number365965
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/15/2019
Supplement Date Received by Manufacturer04/15/2019
Initial Report FDA Received Date04/30/2019
Supplement Report FDA Received Date05/09/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age18 YR
-
-