• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. TWINFIX ULTRA PK 5.5MM W/ 2 UB-BL &BLK; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. TWINFIX ULTRA PK 5.5MM W/ 2 UB-BL &BLK; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number 72202625
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/05/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a procedure the anchor snapped during insertion into bone.No patient injuries reported and no delay.The procedure was successfully completed with the same device.All the pieces from the loose body anchor were retrieved with a arthroscopic grasper.
 
Additional Manufacturer Narrative
One 5.5 twinfix ultra pk anchor assembly was returned for evaluation.Only the proximal end of the anchor was returned, no inserter, sutures or packaging.Visual assessment confirmed the reported breakage.The distal tip of the anchor has broken at the suture eyelet.The break is angular in nature indicating off axis insertion.Dimensional assessment of the anchor major diameter and wall thickness was performed confirming it met print specifications.Without the return of the inserter an exact root cause cannot be determined with confidence; however, factors that could have contributed to the reported event include: use of excessive force during insertion which can cause failure of the suture anchor or insertion device.A two-finger ao technique should be used to insert the anchor.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.This investigation could not identify any evidence of product contribution to the reported complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TWINFIX ULTRA PK 5.5MM W/ 2 UB-BL &BLK
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
MDR Report Key8565787
Report Number1219602-2019-00485
Device Sequence Number11172385
Product Code MBI
UDI-Device Identifier03596010647535
UDI-Public03596010647535
Combination Product (Y/N)N
Initial Reporter CountryEI
PMA/510(K) Number
K093228
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/04/2023
Device Catalogue Number72202625
Device Lot Number2013988
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/25/2019
Initial Date Received by Manufacturer 04/05/2019
Supplement Date Received by Manufacturer05/02/2019
Initial Report FDA Received Date04/30/2019
Supplement Report FDA Received Date05/03/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age65 YR
Patient Weight76
-
-