Model Number V60
Device Problems Failure to Power Up (1476); Therapeutic or Diagnostic Output Failure (3023)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2019.Date of report: 30april2019.A follow-up report will be submitted once the investigation has been complete.
 
Event Description
The customer reported unit does not turn on and blower is not responding.The unit was in clinical use at the time the reported issue was discovered; however, there was no harm to the patient or the user.
 
Manufacturer Narrative
Mfr received date: 22 aug 2019.Date of report received: 29 aug 2019.The assembly,pcb,mtr ctrlr,v60 was returned to failure investigation for evaluation.Visual inspection of the assembly,pcb,mtr ctrlr,v60 revealed no signs of damage or contamination.This customer returned mc (motor controller) board was working correctly.The customer replaced the mc board to address the reported problem.The unit successfully passed the required performance verification test.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key8566582
MDR Text Key143675128
Report Number2031642-2019-02618
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public(01)00884838020054
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 04/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/30/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/24/2019
Was the Report Sent to FDA? No
Date Manufacturer Received04/11/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-