• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT RX HERCULINK ELITE STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT RX HERCULINK ELITE STENT SYSTEM Back to Search Results
Catalog Number 1011499-18
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/11/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The hi-torque spartacore device referenced is being filed under a separate medwatch report number.A visual was performed on the sds.The reported difficulty to remove from the guide wire was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
It was reported that the procedure was to treat a distal renal artery.A hi-torque spartacore guidewire (gw) and a herculink elite stent delivery system (sds) were advanced without issues and the herculink was implanted.However, when removing the sds, resistance was felt with the gw.The sds was pulled out by hand with the guide wire stuck in the sds, as a single unit.There were no adverse patient effects and no clinically significant delay during the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RX HERCULINK ELITE STENT SYSTEM
Common Device Name
STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key8566964
Report Number2024168-2019-03442
Device Sequence Number2293349
Product Code NIN
UDI-Device Identifier08717648078200
UDI-Public08717648078200
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P110001
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date (Section B) 04/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2021
Device Catalogue Number1011499-18
Device Lot Number8110861
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/19/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/11/2019
Initial Report FDA Received Date04/30/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age85 YR
Patient Weight75
-
-