| Model Number |
MMT-523LNAP |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/28/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A test reservoir was installed and did not lock in place due to completely broken reservoir tube lip.Unable to perform the displacement test due to completely broken reservoir tube lip.Insulin pump had missing reservoir tube o-ring.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that their insulin pump was damaged.Customer stated the reservoir lip almost cracked completely.The customer¿s blood glucose level was 190 mg/dl at the time of the call.Customer stated the reservoir still locks into place.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump was returned for analysis.
|
| |
|