STRYKER NEUROVASCULAR CORK NEUROFORM ATLAS STENT SYSTEM 3.0 X 21MM WITHOUT TIP; STENT, INTRACRANIAL NEUROVASCULAR Back to Search Results
Catalog Number M003SZAS30210
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/24/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Subject device is not available.
 
Event or Problem Description
It was reported that during advancement of the stent (subject device) through a microcatheter, the stent prematurely disconnected from the delivery wire within the microcatheter.The stent and microcatheter were removed as one unit.There were no clinical consequences to the patient.
 
Event or Problem Description
It was reported that during advancement of the stent (subject device) through a microcatheter, the stent prematurely disconnected from the delivery wire within the microcatheter.The stent and microcatheter were removed as one unit.There were no clinical consequences to the patient.
 
Additional Manufacturer Narrative
The device history record confirms that the device met all material, assembly and performance specifications.The subject device was not returned for analysis; therefore, physical as well as a functional testing could not be performed.The reported issue is covered in the device directions for use.The risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.Information available indicated that the device was confirmed to be in good condition prior to use.It is probable that there were some procedural/anatomical factors that contributed to the event; this however cannot be conclusively determined.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause, a cause of undeterminable was assigned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROFORM ATLAS STENT SYSTEM 3.0 X 21MM WITHOUT TIP
Common Device Name
STENT, INTRACRANIAL NEUROVASCULAR
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
MDR Report Key8567575
Report Number3008881809-2019-00114
Device Sequence Number14026641
Product Code NJE
UDI-Device Identifier07613252656159
UDI-Public07613252656159
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
H020002/S046
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/04/2023
Device Catalogue NumberM003SZAS30210
Device Lot Number21069100
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/24/2019
Supplement Date Received by Manufacturer05/15/2019
Initial Report FDA Received Date04/30/2019
Supplement Report FDA Received Date06/03/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
XT-17 MICROCATHETER (STRYKER)
Patient Age52 YR
-
-