| Model Number |
ADSR01 |
| Medical Device Problem Code |
Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/04/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was in atrial tachycardia.Undersensing due to events falling within the atrial refractory and blanking period was noted on the implantable pulse generator (ipg).Intermittent far field r-wave oversensing was also noted on the right atrial (ra) lead.The ipg and ra lead remains in use.No patient complications have been reported as a result of this event.
|
| |
|