Udi information (b)(4).Sample was received for evaluation.Images were taken of the ¿as received¿ valve.The position of the cam when valve was received was 110mmh2o.The valve was visually inspected: needle holes in the needle chamber and biological debris were noted.No catheters were returned.The valve was hydrated.The valve was tested for programming and the valve passed the test.The valve was flushed and the valve passed the test no occlusion was noted.The valve was leak and the only leaked from the needle holes over the needle guard.The valve was reflux tested per tm-100313 rev.10 the valve failed the test.The valve was dried.The valve was then pressure tested and the valve failed the test.The valve was dismantled and was examined under microscope at appropriate magnification: biological debris was found on the spring, on the spring pillar, on the ruby ball, and on the seat of ruby ball.The root cause for the pressure issue is due to the biological debris found on the spring, on the spring pillar, on the ruby ball, and on the seat of the ruby ball, the biological debris was not letting the ruby ball sit correctly.Manufacturing records were reviewed and found no anomalies.Based on the results of the investigation, the reported issue is confirmed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
|