This report is being submitted for an additional suspect medical device, this event was previously submitted under manufacturer report numbers: 3008344661-2019-00010 and 3008344661-2019-00030 (and there is no further impact to patient management).An investigation was performed for the customer issue and included a review of complaint text, a search for similar complaints, complaint trending report review, testing of retained kit of lot 91396fn00, and a review of product labeling.Ticket searches determined normal complaint activity for all three lots with this being the only complaint received to date for lots 92145fn00 and 94039fn00.Trending report review determined there are no non-statistical or adverse trends.Lot 91396fn00 was tested using a retained kit and all specifications were met, indicating the lot is preforming acceptably.Data was reviewed for lots 92145fn00 and 94039fn00, and no systemic issue was identified.Labeling was reviewed and found to be adequate.Based on all available information and abbott diagnostics' complaint investigation a product deficiency was not identified.
|