Returned product consisted of a nc quantum apex balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed kinks to the hypotube, inflation lumen, guidewire lumen, and core wire.The hypotube is separated 89.2cm from the hub and appears to have been kinked prior to separation.Microscopic examination revealed no damages.There is blood present in the guidewire lumen and the balloon is still tightly folded.Inspection of the remainder of the device presented no other damage or irregularities.
|