| Model Number |
UNI REVERS HUMERAL RESECTION SLIDE BLOCK |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/12/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was requested/is expected but has not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that during a reverse total shoulder arthroplasty in the left shoulder, the cutting block with boom was unable to advance towards the patient's shoulder.The surgeon performed a free handed humeral head cut and went on with the case as usual.It was also reported that upon further inspection of the device, there is a very small screw near the entrance for the boom that is protruding into the area that contacts the boom.
|
| |
|