MAUDE Adverse Event Report: DEPUY FRANCE SAS - 3003895575 CORAIL2 LAT COXA VARA SIZE 12; CORAIL AMT CEMENTLESS IMPLANT : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY FRANCE SAS - 3003895575 CORAIL2 LAT COXA VARA SIZE 12; CORAIL AMT CEMENTLESS IMPLANT : HIP FEMORAL STEM Back to Search Results
Catalog Number 3L93712
Device Problems Biocompatibility (2886); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Tissue Damage (2104); Test Result (2695); No Code Available (3191)
Event Date 10/21/2013
Event Type  Injury  
Manufacturer Narrative
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.Initial reporter occupation: non-healthcare professional: "attorney".(b)(4).
 
Event Description
Pfs and medical records received.After review of the medical records for mdr reportability, the patient was revised on (b)(6) 2013 for metal on metal disease.Update: ppf alleges pseudotumor, metal wear, metallosis and elevated metal ions.Doi: (b)(6) 2009; dor: (b)(6) 2013; (left hip).
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.No device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.In order to determine if a lot related issue was possible, a worldwide complaint database search was performed.A worldwide complaint database search found no additional related reports against the provided product code/lot code combination.Based on the inability to find any additional related reports against the provided product code/lot code combination it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.The reported event is considered one of the possible complications of joint replacement.Investigational inputs were requested as indicated per internal procedures for this failure mode.The complaint information provided has been reviewed for complaint coding, medical device reporting, and other data required by the complaint system.Follow-up for additional event information, if applicable, was conducted utilizing work instruction wi-7915 appendix a.Without the physical complaint sample(s) associated with this report, it was not possible to determine if the device(s) failed to meet specification(s) at the time it was released for distribution.The device(s) associated with this event were used in the treatment of the patient as prescribed by the presiding surgeon.From the event information received, it was not possible to determine the relationship of the device to the reported event.No information received with this individual complaint indicated that a broader investigation or corrective action was necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORAIL2 LAT COXA VARA SIZE 12
Type of Device
CORAIL AMT CEMENTLESS IMPLANT : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY FRANCE SAS - 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex 69801
FR  69801
MDR Report Key8570708
MDR Text Key143741922
Report Number1818910-2019-92200
Device Sequence Number1
Product Code KWA
UDI-Device Identifier10603295168898
UDI-Public10603295168898
Combination Product (y/n)N
PMA/PMN Number
K042992
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup
Report Date 07/10/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/01/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/31/2016
Device Catalogue Number3L93712
Device Lot Number5008690
Was Device Available for Evaluation? No
Date Manufacturer Received10/31/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age78 YR
Patient Weight88
-
-