| Model Number |
24667 |
| Medical Device Problem Code |
Sensing Intermittently (1558)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/26/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The returned device was evaluated.Visual inspection upon receiving revealed no external damage or defects.Connected masimo o3 sensors to customer module and measurements were obtainable on both channels 1 and 2.Measurements were continuous with cable and connector manipulation.Measurements were acceptable on each channel and comparable to a known good module.No loose cabling, internal circuitry damage or defects were observed.Customer complaint was not duplicated.The module is fully functional.
|
| |
|
Event or Problem Description
|
|
The customer reported the o3 left side measurement are randomly dropping down without clinical reason and port 1 of the module is not picking up and reading "sensor initializing".No patient impact or consequences were reported.
|
| |
|