• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MASIMO - 40 PARKER O3 REGIONAL OXIMETER MODULE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MASIMO - 40 PARKER O3 REGIONAL OXIMETER MODULE Back to Search Results
Model Number 24667
Medical Device Problem Code Sensing Intermittently (1558)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/26/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The returned device was evaluated.Visual inspection upon receiving revealed no external damage or defects.Connected masimo o3 sensors to customer module and measurements were obtainable on both channels 1 and 2.Measurements were continuous with cable and connector manipulation.Measurements were acceptable on each channel and comparable to a known good module.No loose cabling, internal circuitry damage or defects were observed.Customer complaint was not duplicated.The module is fully functional.
 
Event or Problem Description
The customer reported the o3 left side measurement are randomly dropping down without clinical reason and port 1 of the module is not picking up and reading "sensor initializing".No patient impact or consequences were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O3 REGIONAL OXIMETER MODULE
Common Device Name
OXIMETER
Manufacturer (Section D)
MASIMO - 40 PARKER
40 parker
irvine CA 92618 1604
MDR Report Key8571399
Report Number2031172-2019-00228
Device Sequence Number11349427
Product Code MUD
UDI-Device Identifier00843997010771
UDI-Public00843997010771
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K160526
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 04/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number24667
Device Catalogue Number9637
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/10/2019
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/11/2019
Initial Report FDA Received Date05/01/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-