| Model Number |
1884380EM |
| Medical Device Problem Codes |
Material Separation (1562); Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/05/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
A health care provider (hcp) reported that the blade was snapped in half.There was no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
On follow up, it was reported that the event occurred prior to use.They were having trouble getting it to work.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis found that there was a residue consistent with biological contaminants on the device.The inner shaft broke 0.576¿ from the distal face of the inner hub which would have resulted in the reported malfunction.There were striations around the outside diameter of the break point indicating metal on metal contact during use.The break point corresponds to the proximal end of the outer tube in the front hub.There was no damage to the distal tip.When viewed under magnification, there was deformation and indentations of the front hub locking area consistent with aggressive use.There were no bends or signs of a concentricity issues.(b)(4).If information is provided in the future, a supplemental report will be issued.
|
| |
|