• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG, PRECISION LED LIGHT SOURCE - INTERNATIONAL KIT; LIGHT SOURCE, FIBEROPTIC, ROUTINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG, PRECISION LED LIGHT SOURCE - INTERNATIONAL KIT; LIGHT SOURCE, FIBEROPTIC, ROUTINE Back to Search Results
Catalog Number 0220220000I
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/02/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported that there was loss of image.
 
Additional Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: cold light source switched off.Probable root cause: ¿ cables ¿ connectors ¿ digital board ¿ power supply ¿ filter / fuse ¿ ac inlet board ¿ main board ¿ transition board ¿ fiber board ¿ intermittence between transition board and ch connector ¿ software ¿scope ¿light source ¿ camera head ¿ coupler ¿ 1488 & 1588 extension cable ¿ intermittence while using rf devices ¿ problem toggling light source or from camera head ¿ connected monitors ¿ leaking ¿ use error ¿ poor grounding ¿ electromagnetic interference (emi) from rf communication, hf surgical instruments, esd, or power surge ¿ manufacturing/ service nonconformity.The reported failure mode will be monitored for future reoccurrence.
 
Event or Problem Description
It was reported that there was loss of image.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, PRECISION LED LIGHT SOURCE - INTERNATIONAL KIT
Common Device Name
LIGHT SOURCE, FIBEROPTIC, ROUTINE
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key8573544
Report Number0002936485-2019-00172
Device Sequence Number2346035
Product Code FCW
UDI-Device Identifier07613327051322
UDI-Public07613327051322
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K082813
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 08/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0220220000I
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/05/2019
Supplement Date Received by Manufacturer04/05/2019
Initial Report FDA Received Date05/02/2019
Supplement Report FDA Received Date08/15/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-