SMITH & NEPHEW MEDICAL LTD. RENASYS TOUCH DEVICE & POWER SUP; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP Back to Search Results
Catalog Number 66801280
Device Problem Failure to Run on Battery (1466)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/25/2019
Event Type  malfunction  
Event Description
It was reported that the unit is not charging.The pump has been changed.
 
Manufacturer Narrative
We have now concluded our investigation for the complaint received.A review of the associated batch manufacturing records did not identify any issue at the point of manufacture which may have caused or contributed to the issue highlighted by the complainant.In addition it can be confirmed that all finished product specification testing was satisfied at the point of release.The returned device was sent to a smith & nephew service centre for a full product evaluation.The device was analysed by our experts and no failure was found with the device.The device performed as expected without issue.This suggests that the reported failure mode was likely caused from the device overheating.To comply with safety regulations this device will fail to charge if it reaches 45°c.This can be caused by the external temperature, how the device is stored and also the heat produced by the device.The device will still carry out npwt at this temperature but will not charge.The temperature of the device can be reduced by storing it in a cooler place, making sure the device is charged prior to use or locking the screen when charging.Due to the fact that the battery could not be charged, the battery alarm was initiated.This alarm is also a safety feature to remind the user that the device may run dead if it is not connected to a power supply.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENASYS TOUCH DEVICE & POWER SUP
Type of Device
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull
UK 
MDR Report Key8574414
MDR Text Key143856763
Report Number8043484-2019-00304
Device Sequence Number1
Product Code OMP
Combination Product (y/n)N
PMA/PMN Number
K153209
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/02/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number66801280
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/25/2019
Date Manufacturer Received10/18/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-