|
The axium prime pusher was returned for analysis without the implant coil, the instant detacher, the microcatheter, or accessories devices.The axium prime pusher was decontaminated.The pusher was found broken distal to the break indicator with the release wire extending out of the proximal end of the pusher.This is indicative that a manual detachment attempt was performed at this location.The actuator interface, positive load indicator, coupler tube, and break indicator were missing and not returned with the pusher for analysis.Bends were found along the length of the pusher.The coin was found to be present and pulled back from the lumen stop; with the shield coil present and intact.The lumen stop was found to be visually acceptable, however, there was damage found to the inner diameter of the retainer ring.In addition, a part of the detach element was found broken and stuck within the retainer ring.The lumen stop inner diameter (id) was measured to be within specification.It was not possible to take the measurement of the retainer ring inner diameter (id) due to the damage.No other anomalies were observed.Based on the analysis performed, the axium prime coil was confirmed to have non-detachment as the pusher was returned with the detach element stuck within the retainer ring, however, the cause of the damage could not be determined.Based on the returned device, the pusher was found bent in two locations along its length, indicative of either high tortuosity or damage during return shipping to medtronic for analysis.There was no non-conformance to specification identified that led to the non-detachment.The review of lot history records shows that the finished device has met all manufacturing requirements and specifications during final assembly and quality inspection.Since the implant coil, actuator interface, positive load indicator, coupler tube, break indicator and instant detacher were not returned; any contribution of the implant coil, actuator interface, positive load indicator, coupler tube, break indicator and instant detacher to the non-detachment could not be determined.Per the axium coil and i.D.(instant detacher) instructions for use (ifu): ¿if the coil does not detach after 3 attempts, discard the i.D.(instant detacher) and replace with a new i.D.(instant detacher).Also, in the rare event that the coil does not detach and cannot be removed from the implant delivery pusher, use the following steps for detachment.A.Grip the hypotube approximately 5 cm distal of the positive load indicator at the hypotube break indicator and bend the implant delivery pusher just distal to the hbi 180 degrees.B.Next straighten the pusher back, continue bending and straightening until the pusher tubing opens exposing the release element.C.Gently separate the proximal and distal ends of the open pusher.Then, under fluoroscopy, pull the proximal portion of the implant delivery pusher approximately 2-3 cm to confirm implant detachment per ifu.¿ medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|