Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Lot specific voluntary recall was initiated for the lfit v40 cocr heads within scope of a capa.The investigation revealed that only specific catalog numbers and specific lots are impacted by the regulatory action and that the affected lots were manufactured on or before march 4, 2011.The root cause analysis identified a process related anomaly as to the affected sizes and lots.The affected product has all been implanted and/or expired.
|
|
It was reported that the patient's left hip was revised due to trunnionosis.Intra-operatively, excessive trunnion wear, a moderate amount of black tissue in the patient, and liner wear were noted.An accolade tmzf plus stem, lfit v40 head, and 40f liner were revised to a restoration modular stem, metal head, and liner.Rep provided primary implant stickers and reported that no further information will be released by the hospital or surgeon.
|