• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS, INC MICRO-INTRODUCER KITS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS, INC MICRO-INTRODUCER KITS Back to Search Results
Catalog Number H787065970155
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 02/01/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reported defective device has yet to be returned to the manufacturer for a device evaluation.Angiodynamics is attempting to obtain the device.An investigation into the root cause for product problem is currently in progress.The results of the device evaluation.Will be sent via a follow up medwatch.Complaint # (b)(4).
 
Event or Problem Description
Per received user medwatch 0504240000-2019-8006: pt undergoing cerebral angiogram the site was injected with lidocaine 1 % overlying the right common femoral artery the groin and arterial access was gained using a micropuncture technique, with the access wire in place the md went to remove the micropuncture sheath the hub disconnected from the main body of the sheath - when the md pulled on the main body of the sheath, approximately one half of it came off.Approx.2 1/2 cm of the tip was missing and still on the wire, attempts to dissect with a hemostat down the wire to identify and remove - unsuccessful attempt.The left common femoral artery was accessed with ultrasound guidance, a 6f 11 cm sheath was placed and flushed, through this a snare was used to capture the distal end of the access wire and withdraw out the left femoral artery sheath - a 5cm sheath was placed over the wire into the right common femoral artery - a 70 cm kumpe was passed antegrade down the access wire and the tip of the was pushed out the left side of the groin - there was not harm to the patient all was retrieved.It was indicated the defective disposable device is available for return to the manufacturer for evaluation.
 
Additional Manufacturer Narrative
Returned for evaluation was a sheath and a dilator.A visual inspection noted that the sheath was separated from the hub and broken in 2 pieces.The dilator was noted to be separated from the hub.The sample was forwarded to the vendor facility for evaluation.The customer's reported complaint description of the microaccess micro-introducer sheath was fractured is confirmed.Angiodyanmics' supplier was unable to determine a root cause for the reported failure.Potential root causes could have been due to high insertion force, patient anatomy, and procedureal complications.A review of the device history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.A review of the angiodynamics complaint system noted no trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Complaint # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRO-INTRODUCER KITS
Common Device Name
MICRO-INTRODUCER KITS
Manufacturer (Section D)
ANGIODYNAMICS, INC
603 queensbury avenue
queensbury NY 12804
MDR Report Key8575001
Report Number1319211-2019-00037
Device Sequence Number16767914
Product Code DQT
UDI-Device IdentifierH787065970155
UDI-PublicH787065970155
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K051655
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2021
Device Catalogue NumberH787065970155
Device Lot Number5289467
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/03/2019
Supplement Date Received by Manufacturer04/03/2019
Initial Report FDA Received Date05/02/2019
Supplement Report FDA Received Date08/19/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age51 YR
-
-