| Model Number |
TRUETRACK |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/08/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(manufacturer narrative = f, corrected data = f) internal report # (b)(4).Returned meter evaluated with no defect found.Test strips not returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Test strip udi#: (b)(4).Note: manufacturer contacted customer to ensure that the replacement products resolved the initial concern - able to establish contact with customer who indicated that is comfortable with results from the replacement meter.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for erratic blood glucose test results.The director of nursing is calling on behalf of the customer.The customer is concerned with tests results from results obtained of 275 and 128mg/dl.The expected fasting blood glucose test result range is undisclosed.The customer did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product storage location is undisclosed.The test strip lot manufacturer's expiration date is undisclosed and open vial date is undisclosed.The meter memory was reviewed for previous test result history by qc.Result 1: (b)(6) 2019 3:08:00 pm 275 post lunch blood.Result 2: (b)(6) 2019 3:09:00 pm 128 post lunch blood.
|
| |
|